MCP ER300 EMI TEST RECEIVER / emi test receiver

FOB Reference Price:Get Latest Price
$6,000.00 - $7,000.00 / Piece | 1 Piece/Pieces (Min. Order)
Customization:
Customized logo (Min. Order: 5 Pieces)
Report Suspicious Activity
Overview
Quick Details
Place of Origin:
Shanghai, China
Brand Name:
MCP
Model Number:
ER300
Power:
Electronic
Supply Ability
Supply Ability:
50 Piece/Pieces per Month
Packaging & Delivery
Packaging Details
1pcs per carton
Port
Shanghai
Product Description

Features:

.      Frequency range: 1kHz—300MHz/2GHz/3.6GHz

.      Multiple digital detection functions, includingCISPR-AVG      CISPR-RMSQPK

.      Multiple EMI measurement standards

.      Optimal sensitivity - 145dBm

.      Resolution bandwidth 1Hz~3MHz(-3dB),

                                                 200Hz/9kHz/120kHz/1MHz(-6dB)

.      Plenty of measurement options available                                

.      Double RF input protection ensures measurement safety

.      Support ETR measurement analysis software

.      8.4-inch high-brightness TFT LCD display

.      Universal USB and LAN for convenient control

.      The weight does not exceed 6.5kg

Spectrum analysis mode

Frequency range

ER300

1kHz~300MHz

ER2000

1kHz~2GHz

ER3600

1kHz~3.6GHz

 Reading accuracy

±(frequency indication x frequency accuracy+1%span+10%RBW+0.5x[span/(sweep dot-1)]+1Hz)

Internal time base (10MHz)

Aging rate

1ppm/year

Temperature drift     

0.5ppm(15~35)

SSB phase noise (f=500MHz)

Offset 30kHz

-90dBc/Hz

Offset 1MHz

-110dBc/Hz

·                                 Amplitude

Maximum input level

Average continuous power

 +36dBm

Max. DC Input voltage

 50Vdc

Display average noise level

Channel

Frequency

Preamp off

Preamp on

Frequency conversion channel

100kHz~1MHz

≤-100dBm-30*(f/100kHz)dB

≤-120dBm-30*(f/100kHz)dB

1MHz ~ 10MHz

≤-130dBm

≤-150dBm

10MHz ~ 1GHz

≤-135dBm

≤-155dBm

1GHz ~ 3.6GHz

≤-140dBm

≤-148dBm

Low frequency channel

5kHz~10kHz

 

≤-110dBm

10kHz~10MHz

 

≤-125dBm

Uncertainty ofamplitude  measurement(20~30)

Combined amplitude Accuracy(90%)

±1.8dB

Resolution bandwidth(-3dB)

Resolution bandwidth range

1Hz~3MHzsteps

Uncertainty ofresolution bandwidth conversion

1Hz≤RBW≤500kHz

±0.6dB

RBW500kHz

±1.0dB

Resolution bandwidth accurate

10%

Attenuator

Attenuation range

0~39dB3dBsteps

Attenuatoruncertainty

±1.0dB

Residual response

≤ -96dBm

Frequency response

5kHz~200kHz

±1.8dB

200kHz~3.6GHz

±1.5dB

 Detector

Autonormal +peak-peaksampling

Sweep time

Conversion channel

Span = 0

1ms~3000s

Span > 0

3ms~3000s

Input port SWR

50MHz~1GHz

≤ 2.0

EMIreceiving mode

Frequency range

ER300

1kHz~300MHz

ER2000

1kHz~2GHz

ER3600

1kHz~3.6GHz

Reading accuracy

±(frequency indication x frequency accuracy + 0.5 display unit)

Uncertainty of amplitude measurement(20~30)

Combined amplitude accuracy(90%)

±2.0dB

Resolution bandwidth (-6dB)

Resolution bandwidth range

200Hz/9kHz/120kHz/1MHz

uncertainty ofresolution bandwidth

10%

Detector

+peak-peakqoasi peakaverageRMS

Sweep time

100us~100s

Sweep point

101~1001

Trace number

3(parallel detection)

Frequency band list

10

Frequency response

±2.0dB

Tracing source

Frequency range

10kHz~1.5GHz

Output power

-30dBm~0dBm

Output flatness

±3dB

Common parameter

Display

8.4”TFT-LCD

Resolution

800×600

Communication interface

LANUSB

Operating temperature

0~40

Storage temperature

-30~+70

Weight

7.5kg

Dimensions (l * w * h)

400mm×280mm×190mm